Measuring Phase Noise with an X-Series Analyzer

展示

This video demonstrates how to make a phase noise measurement with a Keysight X-Series signal analyzer. It shows how to find a carrier with auto tune, open the phase noise application, read the results, and calculate jitter over a bandwidth you choose.

 

Set up the signal

The demonstration connects a source that transmits a continuous wave (CW) signal at 20 GHz with an amplitude of 0 dBm. In the analyzer's default state, the signal is visible on screen. The auto tune function tunes the analyzer to the signal even when none of its parameters are known. The analyzer finds the signal and places a marker at the peak, which confirms the frequency and amplitude.

 

Open the phase noise application

Open the mode, measurement, and view selector. Select phase noise as the mode, log plot as the measurement, and decade table as the view, then click OK. In the phase noise application, click auto tune again. The analyzer finds the carrier frequency and takes a single phase noise measurement. To repeat the measurement at any time, select restart.

 

Read the phase noise plot

By default, the measurement starts at an offset of 100 Hz above the carrier frequency and stops at an offset of 1 MHz above the carrier frequency. The plot shows two traces. The yellow trace is raw and represents the actual captured values. The blue trace is smoothed, which means it has been averaged. The carrier lies to the left of the traces, so the carrier power is not visible on the plot, but the application displays the carrier power along with the reference value, which is the highest power level visible.

 

Use the decade table and markers

The decade table presents the phase noise levels at offsets of 100 Hz, 1 kHz, 10 kHz, 100 kHz, and 1 MHz. You can also add markers at frequencies of your choice.

 

Calculate jitter

You can calculate jitter across all or part of the measurement bandwidth. Select a marker, go to the marker function, and choose integrated RMS noise from the drop-down. Select the second jitter option as the integrated RMS noise unit, then enter the start and stop frequencies of the measurement bandwidth you want. The jitter value appears on screen.

 

What you will learn

  • How to use auto tune to find a carrier and confirm its frequency and amplitude
  • How to open the phase noise application with a log plot and decade table
  • How to repeat a measurement with restart
  • How to read the raw and smoothed traces and the default 100 Hz to 1 MHz offset range
  • How to use the decade table and markers
  • How to calculate jitter over a bandwidth you choose