Procedure overview

  1. Preset system
  2. Perform compensation of fixture
  3. Set measurement parameters
  4. Connect DUT and measure at a single frequency
  5. Set up list sweep table
  6. Set trigger mode and make sweep measurement

In this demo...

  • Measurement of chip SMD device
  • Fixture compensation
  • List sweep

1. Preset system

a. Press [Preset] front panel key

b. Press CLEAR SET&CORR softkey

c. Press OK softkey (wait about 20 seconds…)

2. Perform compensation of fixture

a. Set fixture to UNKNOWN terminals

b. Press [Meas Setup] front panel key

c. Select CORRECTION softkey

d. Move cursor onto OPEN field by pressing

[▼] front panel key

e. Make open condition for the test fixture