The 8753D firmware upgrade, revision 07.14, is compatible with Windows 3.1, WindowsNT and Windows 95 operating systems1.

Firmware Upgrade Compatibility

  • All 8753D with firmware revision 7.10 or higher.
  • Other firmware versions require the 8753D Option 000, CPU upgrade kit.

Contact your local Keysight service center to schedule this CPU upgrade.

Downloading Instructions

To download the 8753D firmware upgrade, click on the hyperlink provided. When the "Save As" dialog box appears, save the 8753d.exe file to your floppy or hard disk.

This file is an MS-DOS self-extracting file. Select start, run and type in the path to the file, or press the browse button to navigate through the directory structure to the file. Select the OK button. This process will uncompress the 8753D_07._14 file2.

Installation Instructions

Save the 8753D_07._14 file to an MS-DOS formatted 3.5-inch disk. Insert the disk into the network analyzer's floppy-disk drive. If the power is off, turn on the network analyzer to upload the new firmware into the instrument. If the instrument is on, then simply cycle the power to upload the firmware. Once the file has finished uploading, the instrument will display the new firmware revision number. After installation of the firmware upgrade, the 8753D vector network analyzer is ready for use.

1MS-DOS, Windows 3.1, WindowsNT and Windows95 are U.S. registered trademarks of Microsoft Corporation.

2Do not change the filename. The network analyzer will not recognize and load the file if the filename has been changed.

キーサイトは、最新のバグ修正とセキュリティパッチを含む、利用可能な最新バージョンを使用することをお勧めします。
リリース バージョン 内容
2004-02-20 07.14 8753d.exe
リリース バージョン
2004-02-20 07.14
Version Description
8753d.exe

ダウンロード方法

  • See Above

インストール方法

  • See Above

ライセンス

  • Does Not Require a License

Installs on: Instrument