E443xB ESG Signal Generator Option UN8 or UN8/UN9, Software Patch
Situation:
The Baseband Generator board with part number E4400-60070, serial number starting with 300, and version number A0110 contains wrong header information. This condition generates I/Q offset problems. To check if your E443xB ESG-D has the defect, access the installed board information on the display of the ESG. Using the front panel press Utility > Instrument Information > Installed Board Information.
Solution:
To remedy the problem, download and run the self extracting executable Software Patch "BBG_rev.exe" available below.
Keysight는 가능한 최신 버전을 사용하는 것을 권장하며, 최신 버전에는 현재 사용 가능한 가장 최신 버그 수정 및 보안 패치가 포함되어 있습니다.
| 릴리스 날짜 | 버젼 | 버전 설명 |
|---|---|---|
| 2004-02-20 | 1.0 | Self-extracting executable Software Patch. |
| 릴리스 날짜 | 버젼 |
|---|---|
| 2004-02-20 | 1.0 |
| Version Description |
|---|
| Self-extracting executable Software Patch. |
다운로드 방법
- Click the Download button
- Save the file to your PC
설치 방법
- Activate the executable file
- Follow the installation prompts
라이센스
- Does Not Require a License
Installs on: Instrument
선결 조건
None
지원되는 계측기
- E4430B, E4431B, E4432B, E4433B, E4434B, E4435B, E4436B, E4437B with Option UN8 or UN8 UN9.