Situation:

The Baseband Generator board with part number E4400-60070, serial number starting with 300, and version number A0110 contains wrong header information. This condition generates I/Q offset problems. To check if your E443xB ESG-D has the defect, access the installed board information on the display of the ESG. Using the front panel press Utility > Instrument Information > Installed Board Information.

Solution:

To remedy the problem, download and run the self extracting executable Software Patch "BBG_rev.exe" available below.

Keysight empfiehlt die Verwendung der aktuellen Version, die die neuesten verfügbaren Fehlerbehebungen und Sicherheitspatches enthält.
Veröffentlichungsdatum Version Versionsbeschreibung
2004-02-20 1.0 Self-extracting executable Software Patch.
Veröffentlichungsdatum Version
2004-02-20 1.0
Version Description
Self-extracting executable Software Patch.

How to Download

  1. Click the Download button
  2. Save the file to your PC

How to Install

  1. Activate the executable file
  2. Follow the installation prompts

Lizenzierung

  • Does Not Require a License

Installs on: Instrument

Voraussetzungen

  • None

Supported Instruments

  • E4430B, E4431B, E4432B, E4433B, E4434B, E4435B, E4436B, E4437B with Option UN8 or UN8 UN9.
Keine Softwareversion verfügbar
SHOW MORE