Situation:

The Baseband Generator board with part number E4400-60070, serial number starting with 300, and version number A0110 contains wrong header information. This condition generates I/Q offset problems. To check if your E443xB ESG-D has the defect, access the installed board information on the display of the ESG. Using the front panel press Utility > Instrument Information > Installed Board Information.

Solution:

To remedy the problem, download and run the self extracting executable Software Patch "BBG_rev.exe" available below.

A Keysight recomenda usar a versão mais atual, que inclui as correções de bugs e arquivos de segurança mais atualizados disponíveis.
Data de lançamento Versão Descrição da Versão
2004-02-20 1.0 Self-extracting executable Software Patch.
Data de lançamento Versão
2004-02-20 1.0
Version Description
Self-extracting executable Software Patch.

Como Baixar

  1. Click the Download button
  2. Save the file to your PC

Como Instalar

  1. Activate the executable file
  2. Follow the installation prompts

Licenciamento

  • Does Not Require a License

Installs on: Instrument

Pré-requisitos

  • None

Instrumentos Compatíveis

  • E4430B, E4431B, E4432B, E4433B, E4434B, E4435B, E4436B, E4437B with Option UN8 or UN8 UN9.
Não há versão de software disponível
SHOW MORE