Demonstrações
Accurate noise figure measurements start with calibration. This video shows how to calibrate a noise figure measurement on a Keysight X-Series signal analyzer so the result reflects the performance of your device under test (DUT), not your cables, connections, or the analyzer itself.
Why calibrate first
The calibration lets the signal analyzer characterize and remove the effects of cabling and connections, as well as its own contribution to noise. After calibration, the measured noise figure gives an accurate representation of DUT performance.
Connect the calibration setup
Connect the noise source to the dedicated port on the rear of the analyzer. Then connect the noise source RF output directly to the analyzer RF input. Ideally, use two cables with connections that let you insert the DUT between them once calibration is complete.
Select the noise figure measurement
Start or preset the X-Series analyzer, then change the measurement mode and select noise figure. The display shows two plots: noise figure on top and gain below. Two red messages, one at the top and one at the bottom of the screen, show that the measurement is not yet calibrated.
Set the frequency range and points
Enter the measurement frequency range. This demonstration sweeps from 12 GHz to 16 GHz with 21 measurement points. You can also measure at a fixed frequency or create your own list of measurement frequencies within the measurement range.
Configure the DUT and ENR table
Open Measurement Setup, then DUT Setup and Calibration. A schematic of the measurement setup appears, and you can change its parameters. This demonstration measures an amplifier, and you can choose other DUT types. The output IF section is greyed out because an amplifier does not perform frequency conversion. You can also add loss to represent cables or other components. With an SNS noise source, the analyzer reads the excess noise ratio (ENR) data from the source's EEPROM and fills in the ENR table automatically. With a source that has no such memory, you add the ENR data manually.
Run the calibration
Click Calibrate, then click OK to start. When the calibration finishes, the cal state changes to green "cal" and the red messages disappear. After you reduce the scale per division on the Y axis, the traces sit close to 0, which confirms that the measurement is calibrated and ready for use.
What you will learn