FieldFox S-parameter Measurement Uncertainty

데이터 시트

FieldFox S-Parameter Measurement Uncertainty

N9950A

N9951A

N9952A

Corrected Measurement Uncertainty for N9950A/51A/52A

Full 2-port calibration, 85561A 2.92 mm (f) calibration kit, specifications

Power level of -15 dBm, 10 Hz IF bandwidth, no averaging, battery saver off, and 60-minute warm-up time. Includes uncertainties due to drift, noise, compression, and dynamic accuracy. Coverage factor of x1 applied to uncertainties, for ease of comparison with other industry handheld analyzers.

Corrected Measurement Uncertainty for N9950A/51A/52A (continued)

Full 2-port calibration, 85562A 2.92 mm (m) calibration kit, specifications

Power level of -15 dBm, 10 Hz IF bandwidth, no averaging, battery saver off, and 60-minute warm-up time. Includes uncertainties due to drift, noise, compression, and dynamic accuracy. Coverage factor of x1 applied to uncertainties, for ease of comparison with other industry handheld analyzers.

Corrected Measurement Uncertainty for N9950A/51A/52A (continued)

Full 2-port calibration, 85563A 2.4 mm (f) calibration kit, specifications1 Power level of -15 dBm, 10 Hz IF bandwidth, no averaging, battery saver off, and 60-minute warm-up time. Includes uncertainties due to drift, noise, compression, and dynamic accuracy. Coverage factor of x1 applied to uncertainties, for ease of comparison with other industry handheld analyzers.

Corrected Measurement Uncertainty for N9950A/51A/52A

Full 2-port calibration, 85564A 2.4 mm (m) calibration kit, specifications1 Power level of -15 dBm, 10 Hz IF bandwidth, no averaging, battery saver off, and 60-minute warm-up time. Includes uncertainties due to drift, noise, compression, and dynamic accuracy. Coverage factor of x1 applied to uncertainties, for ease of comparison with other industry handheld analyzers.