Highlights

  • Excellent SWR for reducing mismatch uncertainty
  • 50 MHz calibration port
  • Accurate calibration and traceability to US National Institute of Standards and Technology (NIST) millimeter-wave sensor calibration
  • Compatible with EPM (new N1913A/ 14A, E4418B/ 19B), EPM-P (E4416A/ 17A) and P-series (N1911A/ 12A) power meters, plus the E1416A VXI and discontinued 70100A and 43X power meter
  • Accurate average power measurements over -30 to +20 dBm
  • Frequency range 60 GHz to 90 GHz
  • Diode power sensing element

KEY SPECIFICATIONS

Channels
1
Compatible Power Sensors / Power Meters
N1913A/14A and E4418B/19B EPM power meters N1911A/12A and N8262A P-series power meters
Connector Type
Waveguide flange: UG-387/U Flange, EIA WR-12
Features
Maximum SWR: Option 100: <1.06 Option 200: <1.28
Frequency Range
60 GHz to 90 GHz
Maximum Frequency
90 GHz
Channels
Compatible Power Sensors / Power Meters
Connector Type
Features
Frequency Range
Maximum Frequency
1
N1913A/14A and E4418B/19B EPM power meters N1911A/12A and N8262A P-series power meters
Waveguide flange: UG-387/U Flange, EIA WR-12
Maximum SWR: Option 100: <1.06 Option 200: <1.28
60 GHz to 90 GHz
90 GHz
View More
Channels:
1
Compatible Power Sensors / Power Meters:
N1913A/14A and E4418B/19B EPM power meters N1911A/12A and N8262A P-series power meters
Connector Type:
Waveguide flange: UG-387/U Flange, EIA WR-12
Features:
Maximum SWR: Option 100: <1.06 Option 200: <1.28
Frequency Range:
60 GHz to 90 GHz
Maximum Frequency:
90 GHz
Maximum Power:
+20 dBm (100 mW)
Measurement Type (Supported Power Meters):
N1913A/14A and E4418B/19B EPM power meters N1911A/12A and N8262A P-series power meters
Measurement Type:
Average/CW only
Power Range:
Option 100: 1 μW to 100 mW (-30 to +20 dBm) Option 200: 1 nW to 100 mW (-60 to +20 dBm)

Standard Configurations

Interested in a E8486A?

Want help or have questions?