Instructions for Program Update

  1. Left-click on the link "E4991A Option 007 Update" shown below and save the file (CompCalc.exe) to a temporary directory (e.g. C:TEMP) on your computer's hard drive.
  2. Copy the file (CompCalc.exe) to a floppy disk.
  3. Update the program by using the floppy disk and the instruction "E4991A Option 007 Update Procedure" available below.

Known issue for the E4991A Option 007 Temperature Characteristic Test Kit*1 (Fixed)

The temperature drift compensation function of the E4991A option 007 (Temperature Characteristic Test Kit) does not work properly when "Permeability" is selected as the "Material Type" setting*2.

*1: Only the E4991A that has the option 002 (Material measurement firmware) along with the option 007 is affected by this issue. For the E4991A that doesn't have the option 002, the temperature drift compensation function works properly.

*2: When the "Material Type" setting is "Impedance" or "Permittivity", the temperature drift compensation function work properly.

Keysight는 가능한 최신 버전을 사용하는 것을 권장하며, 최신 버전에는 현재 사용 가능한 가장 최신 버그 수정 및 보안 패치가 포함되어 있습니다.
릴리스 날짜 버젼 버전 설명
2009-07-09 1.10 E4991A Option 007 Update
릴리스 날짜 버젼
2009-07-09 1.10
Version Description
E4991A Option 007 Update

다운로드 방법

  • Click the download button and save the file to a temporary directory on your PC (or instrument).

라이센스

  • Does Not Require a License

Installs on: Instrument

운영 체제

  • Windows 2000

선결 조건

  • No specific prerequisites

지원되는 계측기

  • E4991A
사용 가능한 소프트웨어 버전이 없습니다.