HIGHLIGHTS

  • Specified wavelength range: 700 – 1650 nm; operating wavelength range 700 – 1700 nm
  • Absolute wavelength accuracy: ±1.5 ppm (typical ±1 ppm within 15 °C to 35 °C)
  • Scan repetition time less than 0.6 seconds
  • Operating temperature from 0 °C to +55 °C

For the measurement of light sources and signals in wavelength ranges from 700 nm to 1700 nm, the 86120D Multi-Wavelength Meter offers a compelling set of features. With ±1.5 ppm accuracy (typical ±1 ppm) and a measurement cycle time of just 0.6 seconds, it is ideal to achieve fast, automated wavelength adjustment procedures.

Keysight multi-wavelength meters are Michelson interferometer-based instruments that measure wavelength and optical power of laser light over a specified wavelength range. A built-in HeNe laser guarantees an accurate absolute wavelength measurement against a natural standard any time. Simultaneous measurements of multiple laser lines are performed, allowing measurements of DWDM signals and multiple lines of Fabry-Perot lasers. A broadband mode is also available for measuring the weighted center of wider lines.

Benefits:

  • Just 15 minutes of warm-up time are sufficient for specified accuracy.
  • Angled-connector option is available to avoid influencing non-isolated sources.
  • The wide wavelength range covers all communication test applications.
  • With +18 dBm safe input level, optical power values up to +10 dBm can be measured without additional equipment (e.g. attenuator).
  • The built-in applications for optical signal-to-noise ratio, wavelength and power drift allow smart stand-alone test solutions with compact footprint, avoiding costly test procedure development.

This product uses Sumatra PDF on Windows 10-based systems or Foxit Reader for Windows on systems running Windows 7 and older.

For more information about Wavelength Meters, please visit Keysight Wavelength Meters.

Key Specifications

Absolute wavelength accuracy
± 1 ppm
Best Power Accuracy
± 0.6 dB
Broadband Mode
yes
Maximum Linewidth in Narrowband Mode
10 GHz
Measurement Cycle Time
0.6 s
Minimum Resolvable Line Separation
15 GHz
Absolute wavelength accuracy
Best Power Accuracy
Broadband Mode
Maximum Linewidth in Narrowband Mode
Measurement Cycle Time
Minimum Resolvable Line Separation
± 1 ppm
± 0.6 dB
yes
10 GHz
0.6 s
15 GHz
View More
Absolute wavelength accuracy:
± 1 ppm
Best Power Accuracy:
± 0.6 dB
Broadband Mode:
yes
Maximum Linewidth in Narrowband Mode:
10 GHz
Measurement Cycle Time:
0.6 s
Minimum Resolvable Line Separation:
15 GHz
Type:
Multi-Wavelength Meter
Wavelength range:
700 nm to 1650 nm
86120D Multi-Wavelength Meter

Interested in a 86120D?