Overview

This VBA macro performs the 2/3/4 port TRL calibration for any selected port. The TRL calibration provides accuracy equivalent to full 2/3/4/ port calibration using short, open, load, and thru (SOLT) standards.

To perform the TRL calibration, you need to prepare thru, reflection (open or short), and line standards. The thru and line must have the same reference impedance Z0 and the same transfer constant (same material).

Revision Number

A.01.00

Required ENA Firmware Revision

Rev.A.02.xx

Install Procedures

Step1. Download a compressed file from the right side of this page.

Step2. Extract the compressed file by double-clicking it.

Step3. Save all files to the floppy disk.

default.dat

E507X_COMTRL.dll

setup.vbs

SysDefault.dat

TRL_Calibration.vba

Step4. Insert the floppy disk to the floppy disk drive of your ENA.

Step5. Press [Save/Recall] {Explorer} and double-click "A:\setup.vbs"