Important Notes

  • More than two measurement options may not fit in an instrument with prefix < US4207 that originally shipped with only 48 MB of flash memory.  See PSA Series Memory Calculator.
  • Turn off your firewall before starting the PC Update Program, disable or disconnect from any wireless LAN connections, firewall programs interfere with the instrument booting from the PC.
  • If you are using Windows XP, see Using PC Update Program with Windows XP.  
  • Previous versions of firmware do not support hardware and measurement applications released after the firmware version was released.

Having trouble with the update?

For help, start with the Trouble Shooting Wizard

 

 

 

Release Date Version Version Description
2012-07-22 A.11.21 Fixed spike that occurs in SA mode when measuring noise-like signals with signal levels about -50 dBm where Avg/VBW type is set to Log-Power (Video) and the Video bandwidth is wide open (typically >= 4 times RBW). Fixed various Measuring Receiver de
Release Date Version
2012-07-22 A.11.21
Version Description
Fixed spike that occurs in SA mode when measuring noise-like signals with signal levels about -50 dBm where Avg/VBW type is set to Log-Power (Video) and the Video bandwidth is wide open (typically >= 4 times RBW). Fixed various Measuring Receiver de

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA
     

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

  • See Documentation

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Keysight p/n 8121-0545
  • CD-ROM drive
     

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

How to Download

  • Information Not Available

How to Install

  • Information Not Available

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Information Not Available

How to Download

  • Information Not Available

How to Install

  • Information Not Available

Licensing

  • Does Not Require a License

Installs on: PC / Instrument

Prerequisites

  • Information Not Available
SHOW MORE