Official Used
Equipment Store

Oscilloscope Glossary

What Is an Oscilloscope Signal?

Table of contents
Last updated: Nov 25, 2026
Second-Hand. First-Class.

Don’t miss special offers, news, and announcements on the latest used equipment. Unsubscribe at any time.

Why Buy From Keysight

Oscilloscopes

Solve Signal Issues Before They Escalate

See MXR608A Special Offers

An oscilloscope signal is a voltage that changes over time, drawn as a waveform with time along the horizontal axis and voltage along the vertical axis. A digital storage oscilloscope (DSO) captures the signal, stores it and shows it on screen, so you can study the voltage changes.

A triggered scope lines each capture up on one event. With the horizontal delay at zero, where on the screen does that moment land? The obvious answer is the left edge. By default, it isn't.

Overview
  • An oscilloscope signal is a voltage plotted against time. Time runs left to right, voltage runs bottom to top.
  • Common waveforms are square, rectangular, sine and triangle waves.
  • Sample rate and memory depth decide how much detail a digital scope keeps. Sample at least twice the highest frequency in the signal (the Nyquist rate), and expect the scope's own rise time to be about 0.35 ÷ bandwidth (single-pole rule of thumb).
  • The trigger keeps a repeating waveform steady. Volts per division and offset set the vertical range so the signal does not clip.

What Is Waveform and Signal Analysis?

Waveform and signal analysis is the process of reading information out of a waveform. You can do it by eye, or let the scope's measurement functions and analysis software do it for you.

From a waveform you can read the signal's frequency, amplitude and shape.

What Are Square and Rectangular Waves?

Digital circuits produce square and rectangular waves. The voltage sits either high or low, with fast transitions in between. The share of each cycle spent high is the duty cycle.

  1. Square waves spend equal time high and low, so their duty cycle is 50%. Clock signals in digital circuits are usually square waves.
  2. Rectangular waves spend unequal time high and low. Their duty cycle is above or below 50%, so the high pulses are either narrower or wider than the low gaps.
Waveform Shape Typical source What to look at on the scope
Square High and low for equal times, fast edges Digital clocks Duty cycle (50%), rise and fall times
Rectangular High and low for unequal times, fast edges Digital and control circuits, PWM Duty cycle, pulse width
Sine Smooth, repeating curve Oscillators, function generators, AC supplies Amplitude, period, frequency
Triangle Straight rising and falling slopes Function generators Slope, symmetry, peak levels
Waveform
Shape
Typical source
What to look at on the scope
Square
Title
High and low for equal times, fast edges
Title
Digital clocks
Title
Duty cycle (50%), rise and fall times
Rectangular
Title
High and low for unequal times, fast edges
Title
Digital and control circuits, PWM
Title
Duty cycle, pulse width
Sine
Title
Smooth, repeating curve
Title
Oscillators, function generators, AC supplies
Title
Amplitude, period, frequency
Triangle
Title
Straight rising and falling slopes
Title
Function generators
Title
Slope, symmetry, peak levels
Four waveform panels on oscilloscope-style grids: a square wave labelled 50% duty cycle, a rectangular wave with narrow high pulses labelled duty cycle below 50%, a smooth sine wave and a triangle wave with straight rising and falling slopes
Square, rectangular, sine and triangle waves drawn on an oscilloscope-style grid. Source: Keysight illustration.

What Is a Sine Wave?

A sine wave is the smooth, repeating curve that oscillators, function generators and AC supplies produce. A sine wave's amplitude is its peak value, measured from the centre line of the waveform. Its period is the time between two successive peaks, and the frequency is one divided by the period.

On an oscilloscope you read the period in seconds, not a wavelength in meters, because the horizontal axis is time.

A scope shows a signal in time. To see how a filter, cable or amplifier treats a sine wave across many frequencies, you sweep the frequency with a vector network analyzer, which plots results such as insertion loss (see the network analyzer glossary entry). The E5071C (up to 20 GHz, depending on option) is one such analyzer, sold used. Its data sheet carries a discontinuance notice, and Keysight recommends the E5080B for new purchases.

 Keysight MSOX6004A mixed signal oscilloscope showing several waveforms on its screen
A Keysight MSOX6004A mixed signal oscilloscope displays electrical signals as waveforms. Source: Keysight.

What Is a Triangle Wave?

A triangle wave rises and falls in straight lines, so it has sharp peaks and valleys instead of the smooth curve of a sine wave. The rising and falling slopes can be equal or unequal, and its peak-to-peak value is the height from the lowest point to the highest.

How Do Sample Rate and Memory Depth Affect a Signal?

Two settings decide how much of the signal a digital oscilloscope keeps: sample rate and memory depth. Bandwidth matters too: it limits how fast an edge the scope can follow. Keysight's bandwidth glossary gives the rule of thumb rise time ≈ 0.35 ÷ bandwidth for a scope with a single-pole response, so a 500 MHz scope has a rise time of about 0.7 ns.

  • Sample rate is how often the scope measures the voltage, in samples per second (Sa/s).
  • Memory depth is how many samples the scope can store, so it limits how long a stretch of signal you can keep.

The capture window is the memory depth divided by the sample rate. For example, 1 million samples at 1 gigasample per second cover 1 millisecond. A high sample rate and deep memory let you capture a long window and still zoom in on fine detail. To rebuild a signal, the scope must sample at least twice the highest frequency in it, known as the Nyquist rate (Keysight sampling rate glossary).

Diagram of a sine wave with 24 evenly spaced sample points, a row of 24 memory cells holding one sample each labelled memory depth, a bracket between two samples labelled time between samples equals 1 divided by sample rate, and an arrow across the full width labelled capture window equals memory depth divided by sample rate, with the example of 1 million samples at 1 gigasample per second covering 1 millisecond
Each memory cell stores one sample, so the capture window equals the memory depth divided by the sample rate. Source: Keysight illustration.

How Do the Axes and the Trigger Work?

The horizontal (X) axis shows time, in seconds or a fraction of a second. The vertical (Y) axis shows voltage, in volts.

The trigger tells the scope which moment to line every capture up on. The scope waits for a condition you set, such as an edge rising past a chosen voltage level. Because each capture lines up on that same condition, a repeating waveform stays steady instead of drifting. Remember the left-edge guess? Keysight's oscilloscope triggering guide says that with the horizontal delay set to zero, the trigger event sits in the middle of the screen (the default time reference, which you can move), so you also see what happened just before it.

Can Your Oscilloscope Capture Elusive Events?

Elusive events are infrequent, often brief, disturbances that are easy to miss, such as noise bursts, interference and glitches (short unwanted pulses).

To catch them, a scope needs enough sample rate and memory depth to record the event, and a fast waveform update rate. Keysight's white paper on waveform update rate links the chance of catching a rare glitch to how many waveforms the scope draws per second. The page for the InfiniiVision 3000T X-Series lists an update rate of 1,000,000 waveforms per second, so it is one family to compare when you shop for this.

How Does an Oscilloscope Measure a PWM Signal?

Pulse width modulation (PWM) is a digital signal whose duty cycle is varied to control the average power delivered to a device such as a motor or LED. Keysight's measurement help defines duty cycle as the ratio of positive pulse width to the period, so duty cycle is the "on" pulse width divided by the period. A PWM signal with a 50% duty cycle is on for half of each period.

To measure a PWM signal, trigger on the rising edge and read the pulse width and period from the screen.

How Does an Oscilloscope Measure a Sound Signal?

An oscilloscope measures voltage, not sound. To look at sound, use a transducer such as a microphone to turn it into a voltage. The scope then draws that electrical signal as a waveform.

The waveform's amplitude is its height in volts. You can read the peak amplitude or the RMS amplitude. RMS (root mean square) squares the voltage values, averages them and takes the square root. The result is one number that describes the effective voltage of a varying signal (Keysight waves and signals guide).

The frequency is the number of cycles per second, in hertz (Hz). A higher frequency is heard as a higher pitch.

How Do You Use an Oscilloscope to Measure Voltage?

Voltage is the difference in electric potential between two points, measured in volts. A scope input measures the voltage at the probe tip relative to ground, so you connect the probe tip to the point you want to measure and the ground clip to ground. Keysight's probe guide says to connect the ground lead to a known ground. On an Infiniium or InfiniiVision X-Series scope, the channel grounds are tied together and to the electrical mains ground (Keysight support), so Keysight suggests differential active probes to isolate the scope from the circuit's ground. Use a differential probe to measure between two points where neither is ground.

The scope draws the voltage as a waveform, and the height of the waveform on the screen, read against the volts-per-division setting, is the voltage. A 10:1 probe divides the signal by ten and extends the voltage range you can measure. The scope has to know the probe's attenuation to show the right voltage. Many Keysight probes set it automatically through the probe interface (Keysight Infiniium help); for others, set it in the channel's probe menu. Check the probe's maximum input voltage and the scope's input limit in its data sheet before you connect anything.

How Does an Oscilloscope Measure Negative Signals?

A negative voltage is a voltage below the zero-volt reference, which is ground. Whether a signal reads negative depends on which point you call ground.

With DC coupling, the scope shows a negative voltage as a trace below the channel's ground marker, and the readout is a negative number: -5 V sits 5 V below the marker. Check the coupling setting first: AC coupling blocks the steady (DC) part of the signal. The trace is not flipped. Flipping a waveform is a separate vertical inversion control that you switch on yourself.

Why Does Clipping Happen in an Oscilloscope?

Clipping happens when the signal goes beyond the input range of the scope's analog-to-digital converter (ADC). On many scopes the top and bottom of the display match that range, so a clipped waveform looks flattened at the screen edges and no longer shows its true peaks. Keysight's Infiniium help on ADC clipping describes the effect.

A flat-topped trace does not always mean the circuit is faulty. Start with the display settings, then check the circuit.

What you see Possible cause What to try
Flat top or bottom on a waveform Signal exceeds the vertical range Increase volts per division, adjust offset, or use a 10:1 probe
Waveform drifts across the screen Trigger level or source does not match the signal Set the trigger to an edge on the signal and adjust the level
Rough, fuzzy trace Noise or interference Check the ground connection, shorten the ground lead, and look for nearby noise sources
Fine detail missing when you zoom in Sample rate or memory depth too low for the time window Shorten the time window or use a scope with more memory
A rare glitch never appears Too little time spent capturing, or the trigger is not set for it Trigger on the glitch and use a scope with a fast update rate
What you see
Possible cause
What to try
Flat top or bottom on a waveform
Title
Signal exceeds the vertical range
Title
Increase volts per division, adjust offset, or use a 10:1 probe
Waveform drifts across the screen
Title
Trigger level or source does not match the signal
Title
Set the trigger to an edge on the signal and adjust the level
Rough, fuzzy trace
Title
Noise or interference
Title
Check the ground connection, shorten the ground lead, and look for nearby noise sources
Fine detail missing when you zoom in
Title
Sample rate or memory depth too low for the time window
Title
Shorten the time window or use a scope with more memory
A rare glitch never appears
Title
Too little time spent capturing, or the trigger is not set for it
Title
Trigger on the glitch and use a scope with a fast update rate