PSA Series Spectrum Analyzer Firmware

Important Notes

  • More than two measurement options may not fit in an instrument with prefix < US4207 that originally shipped with only 48 MB of flash memory.  See PSA Series Memory Calculator.
  • Turn off your firewall before starting the PC Update Program, disable or disconnect from any wireless LAN connections, firewall programs interfere with the instrument booting from the PC.
  • If you are using Windows XP, see Using PC Update Program with Windows XP.  
  • Previous versions of firmware do not support hardware and measurement applications released after the firmware version was released.

Having trouble with the update?

For help, start with the Trouble Shooting Wizard

 

 

 

Release Date Version Version Description
2011-10-25 A.11.20 Fixed crash when using demodulation measurements with center frequency below 50 kHz. Fixed defect where the RF Power measurement upper frequency was sometimes limited to 3 GHz instead of the upper frequency of the analyzer...

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2011-01-26 A.11.18 Enhanced noise correction algorithm (Noise Correction must be turned on). Fixed issue where displayed PHot and Pcold values were incorrect whenever using external DUT loss.

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2010-02-10 A.11.16 Fixed issue where error message was shown when `Ext LO Frequency' chosen is lower in frequency than the IF Frequency...

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2009-11-11 A.11.13 Fixed issue where VBW key was disabled when switching between RBW to IBW method while in single sweep. Fixed issue where vertical offset lines where not drawn if measurement was set up in single sweep and then the span was changed right before initia

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2009-06-30 A.11.11 Incorporated option 23A functionality (AM/FM/PM triggering) into option 233 as standard. Fixed issue to allow user to set amplitude threshold value for Frequency Counter measurement. Increase timeout for power meter zero and cal to accommodate N5532B

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA
  6. PSA Series Spectrum Analyzer Firmware Revision History

How to Install

Licensing

  • Does Not Require a License
Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2009-01-22 A.11.07 Fixed EVM degradation introduced in A.11.xx firmware when input signal is only transmitted in every other slot. Fixed issue where Low Pass filter setting of None actually used a narrow setting instead of the widest setting. Fixed lock up in Frequency

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2008-10-03 A.11.05 Fixed switch point for phase noise dual loop to single switch point from 1.99 MHz to 2.0 MHz. This improves PSA’s phase noise performance in compatibility mode at a 2.0 MHz span. Fixed modulation accuracy issue where a C8 code channel was inadverte

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2008-07-07 A.11.04 Fixed display regression in PSA firmware A.10.00 through A.11.01 where reference level offset setting would not adjust reference level, although offset was correctly applied to trace data. Fixed issue where SCPI command "corr:cset4:x:space log" would

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2008-05-16 A.11.01 Fixed display issue where pressing the View button while using Max Hold causes trace spikes. Fixed issue where help text for RF burst trigger key was incorrect. Fixed issue where if you set the number of trace points to 7 and use QPD detector, the an

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2007-12-20 A.10.08 Improved preselector centering near the band breaks. Fixed crash related to Amplitude Corrections. Fixed issue where relative level offset was not accounted for when switching between log and linear display. Fixed issue where saved screen image could

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2007-10-31 A.10.07 The EMI Peak detector setup was updated for improved performance with the N9039A RF Preselector and to comply with the current CISPR 16-1-1:2006 standard. Fixed issue where the "CAL:ALL" SCPI query was not properly reflecting the alignment status

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2007-09-24 A.10.05 Fixed regression in A.10.04 firmware that prevented delta markers from working via the front panel.

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2007-03-03 A.09.19 CDMA Mod Accuracy Uplink: Fixed crash. Fixed pilot offset difference between A.08.xx firmware and A.09.09 through A.09.18 firmware.Fixed issue where PVT can not measure idle slot with high noise data channels because slot identification fails...

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2006-12-13 A.09.18 The alignment selection of "off" now functions the same as "alert". Fixed issue where Max Hold glitches when you change frequency with Max Hold on. Fixed crash when mode switching between SA and any Comms app. This crash occurred after about

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2006-10-10 A.09.14 Speed up of Tuned RF Level measurement by a factor of 3 times. Added ability to manually set frequency for Tuned RF Level Measurement. Fixed minor drift issue for E4440A, E4443A, E4445A when making Tuned RF Level relative measurements.

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2006-08-25 A.09.13 Fixed lockup when presetting analyzer and doing a mode switch sequentially. Fixed issue regarding LAN settings being grayed out and not settable after upgrading to A.09 firmware. For option 117 (Secure Memory Erase) instruments, fixed issue causing â

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2006-07-20 A.09.12 Option HB6 (Password Protected Firmware Upgrades) - Updated option from a "bit flip" option to a standard license key enabled option. Note: option HB6, HS1 and HS7 are no longer recommended. New customers are recommended to use option 117...

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2006-06-05 A.09.10 Fixed SRQ Service Request over LAN caused firmware crash when using NI Visa. Fixed Sweep time incorrectly reported when using EMI peak detector. Fixed firmware crash issue by limiting the trace points to 10,000...

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Related Links

2005-11-22 A.08.09 RBW/Span ratio auto coupling now selects correct default for 8560 series on preset. Fixed issue when selecting to use quasi peak detector front panel IF bandwidth limitation changed instrument to inadvertently use base PSA bandwidth values. Command s

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2005-09-02 A.08.07 SA Mode SEM mask regression introduced in version A.08.06 firmware.

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2005-07-27 A.08.06 Options 122 and 140 are now available for the E4443A and E4445A. On-line-help, Trace math, Normalization, and Trace Importation. Correct CISPRAuto-RBW in span 0 Hz now selected. 321.4 MHz Key Label Changed...

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA
     

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2005-03-01 A.07.13 Added Support for option 215 External Source Control. Improved 10 Hz residual sidebands for RBW ≤7 Hz. Fixed Instrument crash during long repeated Spurious Emissions measurements. Fixed Intermittent alignment failure on instrument bootup.

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2004-12-16 A.07.08 Fixed PI/4 DQPSK Constellation Rotates several degrees forward and reverse in phase. Fixed EDGE filter selections does not work in some sequences. Fixed ORFS Switching for MS gives FAIL flag but no offset are failures...

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2004-09-02 A.06.05 Fixed IP forwarding that was not properly configured and could cause network issues.

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2004-08-18 A.06.04 Fixed preselector peaking in bands 5 and 6. Fixed Marker delta amplitude function not working correctly in Basic mode. Fixed Incorrect image saved to file if trace screen is stale or overwritten. Fixed MODE key choices disappear when battery dies.

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2004-03-19 A.05.07 E4446 and E4448 only. Above 26.4 GHz the preselector may not center properly issue fixed.

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2003-02-28 A.04.07 Phase Noise improvement for 1st LO. Fixed if the P-Scramble code is set to > 7, the SSCH Power result is not correct. Fixed trace misalignment of 1 symbol with respect to GSM PFER when Burst Align=1/2 bit offset. Fixed intermittent mask failures for

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2002-10-24 A.03.05 Fixed code domain power quad view channel power measurements results may be in error up to 0.45 dB.

How to Download

  1. Select the firmware version
  2. Select "PSA Firmware Procedure" to guide you through the process
  3. Download the firmware update package into your PC
  4. Install new firmware update package on your PC
  5. Download the firmware from the PC into the PSA

How to Install

Licensing

  • Does Not Require a License

Supporting Documentation

Download

Installs on:

PC / Instrument

Prerequisites

  • Microsoft® Windows based Personal Computer with ethernet wired LAN card.
  • Windows 2000, or Windows XP (Will not work with Windows Vista or Windows 7)
  • Crossover cable if PC or analyzer is not connected to network. Use a CAT, RJ45 cable with cross pinning, Agilent p/n 8121-0545
  • CD-ROM drive
     
  • N9039A Software vs. PSA Firmware Revision Compatibility

Supported Instruments

  • E4440A, E4443A, E4445A, E4446A, E4447A, E4448A, N5531S, N9039A

Related Links

2002-07-08 A.03.04.009 Spectrum Analyzer Firmware upgrade without option 266, revision A.03.04 Release 009.

How to Download

  • Information Not Available

How to Install

  • Information Not Available

Licensing

  • Does Not Require a License
2002-05-30 A.03.03.007 PSA firmware upgrade without option 266, revision A.03.03 Release 007.

How to Download

  • Information Not Available

How to Install

  • Information Not Available

Licensing

  • Does Not Require a License