My Keysight LXI based SICL / LAN instrument is timing out on the LAN connection under automation control. Is it possible to avoid this LAN timeout?

Please access the LXI based instruments web server by entering the instruments IP address into a web browser. If the LAN has timed out it maybe necessary to cycle the power button.
Once the instrument’s internal web page has opened locate the settings pertaining to the TCP Keep Alive as well as the TCP Keep Alive Time.
 

LXI_Keep_Alive_1.png

Alter the TCP Keep Alive and TCP Keep Alive Time as noted and adjust the time as required.

LXI_Keep_Alive_2.png