How does one understand the "Count" in InfiniiVision oscilloscope measurement statistics?

"Count" refers to the waveform count from the time Statistics are enabled or from the time Reset Statistics was pushed. However, this count does not take into account every waveform that has been triggered. Usually, the triggering is faster than the measurements. Therefore, there will be some triggered waveforms that will not be counted in "Count".