How do I convert a percentage(linear) to a dB(logarithmic) value as most of the accuracy values stated in power meter data sheet are in percentage?
1% is equal to 0.0432dB

FAQ 2011-09-12

How do I convert dBm to dBmV since power meter display unit is in dBm?
dBmV = dBm + 47, if the load impedance is 50 Ohm

FAQ 2011-09-12

What is the output update rate of the Recorder Output?
Typically it is about 45 ms.

FAQ 2011-03-31

If I turn OFF the power meter, then turn in ON will the previous zero/cal information still be good?
Some power meters will retain their zero/cal information, some will not….

FAQ 2010-12-20

How do I set the Rise and Fall Times reference level on my N1911A/N1912A P-Series Power Meter?
To set the Rise and Fall Times reverence level on your N1911A or N1912A P-Series Power Meter, you need the firmware A.02.01 or above….

FAQ 2008-09-25