How wide range of resistivity can the 4339B measure when used with the 16008B?
The 4339B can measure surface resistivity up to 7 x 1017 Ω and volume resistivity up to 8 x 1018 Ω-cm, but it depends on the electrodes of the 16008B used.

FAQ 2010-07-02

What are the frequency limitations on the various probes utilized with the Keysight 85070E Dielectric Probe Kit and why?
The 85070E Dielectric Probe options 020, 030 and 050 have maximum frequeny via the E4991A Impedance Analyzer of 3 GHz.

FAQ 2009-06-26

What is the electrical length value defined for the 16454A?
The electrical length of the 16454A has been preset to the E4991A as 0 mm.

FAQ 2008-05-20

Can I use somehow the 16453A for measurements up to 3 GHz?
No, the maximum usable frequency is 1 GHz because measurement uncertainty due to residual impedance of the 16453A rapidly increases at higher frequencies.

FAQ 2008-05-20

Can the 16453A measure film materials thinner than 0.3 mm?
Measurement error due to air-gap significantly increases for thinner MUTs than 0.3 mm. Pre-forming electrodes on both sides of MUT can reduce the error.

FAQ 2008-05-20

Why is it recommended to pre-form electrodes on both sides of thin film materials measured with the 16451B?
Pre-forming electrodes on both sides of film materials can minimize measurement errors due to air gaps between 16451B's electrodes and the material.

FAQ 2008-05-20

What are significant error sources when a thin film material is measured with the 16451B?
Small air gaps become primary error source for the permittivity measurement of thin materials.

FAQ 2008-05-20

Sampling Rate and 44715A Frequency Measurements

FAQ 2002-06-27

44704A Input Filter Specs

FAQ 2002-06-27

Keysight 3852A Rear Panel BNC Drive Specifications

FAQ 2002-06-27

44701A with 44705F or 44708F, Accuracy Specifications

FAQ 2002-06-27

Scan Speed Using the 44702A and the 44710A

FAQ 2002-06-27

Testing the SPEC of 44704A/44711B

FAQ 2002-06-27

Multiple 44702 DVMs Connected to One Computer

FAQ 2002-06-27

Accuracy With the Keysight 3852A and an External DMM

FAQ 2002-06-27

Maximum Number of Waveform Points for the 44726A

FAQ 2002-06-27