What is the life time of the real time clock battery (1420-3056)?
1 year - should be replaced at the instrument's calibration interval.

FAQ 2015-12-14

Can the counter make simultaneous measurements on channels 1 and 2?
No. The channels are independent and can be configured separately; however, they do not have separate processors.

FAQ 2015-09-21

Why will the counter not make low frequency measurements?
Verify the System Timeout setting by pressing Utility - Instr Setup - Time Out Settings. Turn the time out off or set it to a large value.

FAQ 2015-09-01

Why does the counter display Error +580 "No Valid External Timebase"?
53200A counters check for an external time base during the measurement.

FAQ 2015-09-01

My 53200A series counter does not have GPIB. How can I order it?
Contact the Keysight Technical Contact Center and order 53200U-400. This is a $0 upgrade kit which is customer or service center installed.

FAQ 2015-09-01

既に販売が完了したKeysight、AgilentおよびHewlett-Packard(HP)のテスト機器のマニュアルはどのように入手できますか?
ほとんどのKeysight製品にPDFマニュアルがあり、製品ページで入手することができます。Keysightのベンダも古いマニュアルを提供しています。

FAQ 2014-08-01

GP-IB インタフェースだけしかない色々なキーサイト製のアナライザ の測定結果をプリンタに印刷する以外に別の手段はありませんか?
Keysight Intuilink は多くの汎用計測器に接続性を提供する無料で提供されるキーサイトのアプリケーションソフトです。

FAQ 2014-02-08

How do I upgrade my 53200 Series Frequency Counter?
Find Information on Hardware Upgrades, Instrument Software, etc....

FAQ 2013-05-23

53131A/53132A/53181Aおよび53200シリーズのカウンタにオプションを後付けできますか?
この表をご覧になれば、インストール可能なオプションやサービス・センタでのインストールが必要なオプションがわかります。

FAQ 2013-02-12

Keysight 531xxA/53200シリーズ周波数カウンタの入力はフローティングされていますか?それともグランドに接続されていますか?
カウンタの入力はフローティングされていません。BNCのシールド(外側)がシャーシ・グランド(アース)に接続されています。

FAQ 2013-02-12

53131A/53132A/53181A周波数カウンタと通信する場合の、RS-232Cの設定方法は?
RS-232Cで通信できるカウンタはありません。周波数カウンタとPCで通信する場合はGPIBを使用する必要があります。RS-232Cはプリンタ出力にのみ使用します。

FAQ 2013-02-12

How Do I Find My Host Information?
The Host ID is the model and/or serial number of the product, both of which can be found on the serial number label on the back of the instrument.

FAQ 2012-06-14

FAQ:新しいRF/ユニバーサル周波数カウンタへの移行
53131A/53132A/53181Aから53200シリーズRF/ユニバーサル周波数カウンタ/タイマへの移行

FAQ 2011-06-12

What power sensors are available for use with 53147A/53148A/53149A frequency counters?
This FAQ summarizes several questions regarding the use of power sensors and 5314xA counters.

FAQ 2010-11-30

N8480シリーズ・パワー・センサが(標準、CFTオプションのどちらも)使用できない理由は?
435/6/7/8A/Bパワー・メータ、53147/8/9Aマイクロ波カウンタ/パワー・メータ/DVM、70100A MMSパワー・メータ、E1416A VXIパワー・メータ、E7495A/B基地局テスト・セットで、N8480シリーズ・パワー・センサが使用できない理由は...

FAQ 2010-10-31

Why won't Help (.hlp) files view properly with Windows Vista?
In order to view 32-bit Help (.hlp) files with Windows® Vista, you must download the WinHlp32.exe Help program from the Microsoft web site at http://go.microsoft.com/fwlink/?LinkID=82148.

FAQ 2010-10-12

How do I rack mount two instruments of different depths?
There are three ways to rack instruments of different depths side-by-side; for example, a Keysight 33220A with a Keysight 34401A.

FAQ 2010-10-12

53200 Series: How do you set the measurement timeout on the 53210A, 53220A, 53230A counters?
The measurement time-out is the time allowed for each measurement to complete...

FAQ 2010-10-12

周波数カウンタには10 MHz外部タイムベース・オプションが標準で付属しますか?
KeysightのRF/ユニバーサル周波数カウンタ(53181A、53131A、53132A)には、10 MHz内部タイムベースが標準で付属します。

FAQ 2010-09-27

53131A/53132A:タイム・インターバル測定の仕様には、測定結果のレンジに負の値がリストされています。なぜですか?
第2チャネルの終わりが第1チャネルのエッジをわずかに超えると、カウンタは負の値(-1 nsまで)を表示します。差が大きくなりすぎると、カウンタは正の値に切り替わります。

FAQ 2010-09-27

電源をオフした後も、カウンタのファンはなぜ動作し続けるのですか?
内蔵タイムベース(10 MHz発振器)の安定を維持するためです。

FAQ 2010-09-27

53131A/53132A/53181A:53131A、53132A、53181Aカウンタが測定を終了したことはどうしたらわかりますか?
カウンタが測定を終了したことを知るには、次の3つの方法があります。どの方法も、カウンタの標準イベント・ステータス・レジスタを参照して測定の終了を判定します。

FAQ 2010-09-15

カウンタ53131A、53132A、53181Aの入力について
53131A、53132A、53181Aの入力は、フローティングされていません。

FAQ 2010-05-12

Are the N8480 Series Power Sensors compatible with the 53148A counter/power meter/DVM?
No, the N8480 Series Power Sensors are not compatible with the 53148A counter/power meter/DVM.

FAQ 2009-02-04

53131A/53132A/53181A:信号が印加されていないときに、カウンタがGPIBバスをハングアップするのを防ぐ方法は?
カウンタが測定を実行するようにプログラムされている場合は、カウンタは信号が印加されるまで待ってから測定を開始します...

FAQ 2008-04-22