The 83480A operating sytem firmware upgrades are in self-extracting archive files. They MUST be extracted before loading into the instrument.

To download firmware

  1. Create a temporary directory on your computer hard drive.
  2. Click on firmware title listed below. Save it to the temporary directory.
  3. Extract by selecting the temporary directory and unzip the .exe file by double-clicking on the file, or use Run to extract the file (hp834xx.cmp).
  4. Executable contains two files: hp834xx.cmp (firmware upgrade) and readme (explanation of upgrade).
  5. Format a 1.4 Mbyte 3.5" floppy disk in DOS format. It is best to format the disk in the Keysight 83480A mainframe.
  6. Important - label disk with model number and firmware revision. All firmware revisions have the same file name (hp834xx.cmp).
  7. Insert the DOS formated 3.5" floppy disk in you computer disk drive.
  8. Copy the extracted file (hp834xx.cmp) from the temporary directory to the 3.5" floppy.

To upgrade the Keysight 83480A operating system firmware

  1. Press Utility/System config.../Update system firmware/Continue.
  2. It will take approximately 7 minutes to complete the upgrade.

Notes

  • Updating the system firmware will erase any waveforms or states stored in memory.
  • Updating the system firmware will erase the plug-in vertical calibration. Perform a plug-in calibration after upgrading system firmware.
  • Updating the system firmware will not affect the mainframe calibration status.
  • Updating the system firmware will not affect factory optical calibration of the modules.
Keysight는 가능한 최신 버전을 사용하는 것을 권장하며, 최신 버전에는 현재 사용 가능한 가장 최신 버그 수정 및 보안 패치가 포함되어 있습니다.
릴리스 날짜 버젼 버전 설명
2004-04-19 Rev7_12 83480A_Rev7_12.exe
릴리스 날짜 버젼
2004-04-19 Rev7_12
Version Description
83480A_Rev7_12.exe

다운로드 방법

  • See Above

설치 방법

  • See Above

라이센스

  • Does Not Require a License

Installs on: Instrument

선결 조건

  • Not Applicable

지원되는 계측기

  • See Above
사용 가능한 소프트웨어 버전이 없습니다.
SHOW MORE