34460A/34461A/34465A/34470Aファームウェア
34460A/34461A/34465A/34470A DMMのアップデートは、Keysightファームウェア・アップデート・ユーティリティーを使用して、測定器のLAN/GPIB/USB IOポートまたはフロントパネルのUSB「ホスト」ポートから実行できます。
リリース | バージョン | 内容 |
---|---|---|
2021-08-25 | 3.03 | Multiple fixes on Web UI Interactive IO. |
リリース | バージョン |
---|---|
2021-08-25 | 3.03 |
内容 |
---|
Multiple fixes on Web UI Interactive IO. |
ダウンロード方法
Click the Download button.
Save the file to your computer and note the directory where the file is saved. A firmware update installed over the LAN, GPIB, or USB interface is a file with a .upd extension.
Unzip the file to access the .upd file. For updates using a front panel USB drive, the utility extracts the required files from the (unzipped) .upd file.
インストール方法
From the utility directory or Start menu, start the utility (FWUpdate.exe). The utility will prompt you for the location of the .upd file. Follow the additional prompts to complete the installation. The update may take several seconds to complete and will reboot the instrument. After the update is complete, exit the utility.
It is not necessary to re-calibrate the instrument following a firmware-update; however, the instrument may need to be re-secured if it was unsecured to perform the update.
ライセンス
- Does Not Require a License
説明資料
アドオン・ソフトウェア/ユーティリティ
Installs on: Instrument
OS
- Windows 10
前提知識
* NOTICE: Downgrading the 3446xA and 34470A Firmware
Before downgrading to a ‘Previous’ firmware revision, note the firmware revision currently installed (Shift - Help – About).
Firmware revision A.03.01-02.40 and below can be downgraded to any previous (posted) revision.
Firmware revision A.03.01-03.15 cannot be downgraded, and revisions above that can only be downgraded to A.03.01-03.15.* NOTICE: Upgrading the 34460A and 34461A with firmware revision 1.xx to revision 3.0 or greater
34460A and 34461A DMMs with firmware revision 1.08, 1.09, or 1.10 currently installed must first be upgraded to revision 2.17 (see the ‘Previous Versions’ tab), and then upgraded to revision 3.0 or greater. Upgrades to 3.0 or greater (from 1.xx) cannot be done directly. If attempted, the DMM can be recovered by updating the firmware (to 2.17) from the front panel USB port. Select ‘Front Panel’ on the Firmware Update Utility to create the appropriate folder on a USB thumb drive.
対象測定器
- 34460A, 34461A, 34465A, 34470A
ダウンロード方法
Click the Download button.
Save the file to your computer and note the directory where the file is saved. A firmware update installed over the LAN, GPIB, or USB interface is a file with a .upd extension.
Unzip the file to access the .upd file. For updates using a front panel USB drive, the utility extracts the required files from the (unzipped) .upd file.
インストール方法
From the utility directory or Start menu, start the utility (FWUpdate.exe). The utility will prompt you for the location of the .upd file. Follow the additional prompts to complete the installation. The update may take several seconds to complete and will reboot the instrument. After the update is complete, exit the utility.
It is not necessary to re-calibrate the instrument following a firmware-update; however, the instrument may need to be re-secured if it was unsecured to perform the update.
ライセンス
- ライセンス不要
説明資料
アドオン・ソフトウェア/ユーティリティ
インストール先: Instrument
OS
- Windows 8
- Windows 7
- Windows Vista
- Windows XP
- Windows 10
前提知識
* NOTICE: Downgrading the 3446xA and 34470A Firmware
Before downgrading to a ‘Previous’ firmware revision, note the firmware revision currently installed (Shift - Help – About).
Firmware revision A.03.01-02.40 and below can be downgraded to any previous (posted) revision.
Firmware revision A.03.01-03.15 cannot be downgraded, and revisions above that can only be downgraded to A.03.01-03.15.* NOTICE: Upgrading the 34460A and 34461A with firmware revision 1.xx to revision 3.0 or greater
34460A and 34461A DMMs with firmware revision 1.08, 1.09, or 1.10 currently installed must first be upgraded to revision 2.17 (see the ‘Previous Versions’ tab), and then upgraded to revision 3.0 or greater. Upgrades to 3.0 or greater (from 1.xx) cannot be done directly. If attempted, the DMM can be recovered by updating the firmware (to 2.17) from the front panel USB port. Select ‘Front Panel’ on the Firmware Update Utility to create the appropriate folder on a USB thumb drive.
対象測定器
- 34460A, 34461A, 34465A, 34470A
ダウンロード方法
Click the Download button.
Save the file to your computer and note the directory where the file is saved. A firmware update installed over the LAN, GPIB, or USB interface is a file with a .upd extension.
Unzip the file to access the .upd file. For updates using a front panel USB drive, the utility extracts the required files from the (unzipped) .upd file.
インストール方法
From the utility directory or Start menu, start the utility (FWUpdate.exe). The utility will prompt you for the location of the .upd file. Follow the additional prompts to complete the installation. The update may take several seconds to complete and will reboot the instrument. After the update is complete, exit the utility.
It is not necessary to re-calibrate the instrument following a firmware-update; however, the instrument may need to be re-secured if it was unsecured to perform the update.
ライセンス
- ライセンス不要
説明資料
アドオン・ソフトウェア/ユーティリティ
インストール先: Instrument
OS
- Windows 8
- Windows 7
- Windows Vista
- Windows XP
- Windows 10
前提知識
* NOTICE: Downgrading the 3446xA and 34470A Firmware
Before downgrading to a ‘Previous’ firmware revision, note the firmware revision currently installed (Shift - Help – About).
Firmware revision A.03.01-02.40 and below can be downgraded to any previous (posted) revision.
Firmware revision A.03.01-03.15 cannot be downgraded, and revisions above that can only be downgraded to A.03.01-03.15.* NOTICE: Upgrading the 34460A and 34461A with firmware revision 1.xx to revision 3.0 or greater
34460A and 34461A DMMs with firmware revision 1.08, 1.09, or 1.10 currently installed must first be upgraded to revision 2.17 (see the ‘Previous Versions’ tab), and then upgraded to revision 3.0 or greater. Upgrades to 3.0 or greater (from 1.xx) cannot be done directly. If attempted, the DMM can be recovered by updating the firmware (to 2.17) from the front panel USB port. Select ‘Front Panel’ on the Firmware Update Utility to create the appropriate folder on a USB thumb drive.
対象測定器
- 34460A, 34461A, 34465A, 34470A
ダウンロード方法
Click the Download button.
Save the file to your computer and note the directory where the file is saved. A firmware update installed over the LAN, GPIB, or USB interface is a file with a .upd extension.
Unzip the file to access the .upd file. For updates using a front panel USB drive, the utility extracts the required files from the (unzipped) .upd file.
インストール方法
From the utility directory or Start menu, start the utility (FWUpdate.exe). The utility will prompt you for the location of the .upd file. Follow the additional prompts to complete the installation. The update may take several seconds to complete and will reboot the instrument. After the update is complete, exit the utility.
It is not necessary to re-calibrate the instrument following a firmware-update; however, the instrument may need to be re-secured if it was unsecured to perform the update.
ライセンス
- ライセンス不要
説明資料
アドオン・ソフトウェア/ユーティリティ
インストール先: Instrument
OS
- Windows 8
- Windows 7
- Windows Vista
- Windows XP
- Windows 10
前提知識
Prior to updating the firmware, 34460A/34461A/34465A/34470A DMMs with Security Option SEC must have the security feature unlocked. To unlock instrument security, press:
Shift − Utility − Test/Admin – Security – Secure Code
and enter the security code. The security feature can also be unlocked using the command:
CALibration:SECurity:STATe OFF, security_code
The factory-set security code is 'AT3446XA' For additional security this code may have been changed from its factory setting.