When performing materials measurements using the Keysight E4991A and 16453A fixture I obtain unstable measurement results above 1.0 GHz. Why?
Operational "How to" Guides
Summary
When performing materials measurements using the Keysight E4991A and 16453A fixture I obtain unstable measurement results above 1.0 GHz. Why?Description
The Keysight E4991A is specified to 3.0 GHz. However, for material measurements, the Keysight 16453A test fixture is specified to 1.0 GHz. The Keysight 16453A test fixture is the 1.0 GHz limiting factor.
For materials measurements above 1 GHz to best suit your application please see our full line of Materials Test Equipment