What does UNCOR mean on the 71400C display?
UNCOR appears at the upper right of the graticule if your analyzer needs electrical calibration with the built-in correction routine. Since the calibration routine lasts about three minutes, execute it only when the analyzer requires calibration:
- At the beginning of each day
- After extreme temperature changes
- After selecting a new input port
- After moving the analyzer
To execute the electrical calibration routine, follow this about five-minute procedure:
- Connect between 70810B RF Input and 70900B Calibrator
- Push Menu -> Amptd -> CAL ALL