Starting Download..
save Save

Digital Channel Failures on MSO6000 Series and MSO7000 Series Oscilloscopes

There have been some reports of digital channel failures on MSO6000 Series and MSO7000 Series oscilloscopes. A possible failure mechanism has been traced to the firmware on the oscillocopes, particularly system firmware 1.30. The solution for this failure is to upgrade to the latest firmware: InfiniiVision 5000/6000/7000 Series Oscilloscope Software