Starting Download..
save Save

My Keysight 4284A / 4285A LCR Meter is Indicating an "ADC ERR" Error Message. What are the Corrective Actions?

Troubleshooting

Summary

4284A/4285A Error Message Explanation

Question

My Keysight 4284A / 4285A LCR Meter is indicating an "ADC ERR" error message. What are the corrective actions?

Answer

The error message is an indication that the LCR meter's A/D converter within the measurement circuit is malfunctioning. The "ADC ERR" error is a hardware failure.
For additional details please refer to Chapter 3 of the '4285A Precision LCR Meter Operation Manual '.

The "ADC ERR" failure requires service center level repair. Please contact the nearest Keysight Service Center for repair arrangement.

Was this helpful?


Didn't find what you're looking for?