What is the frequency range of the N5532A Sensor Module?
100 kHz to 4.2 GHz, 10 MHz to 18 GHz, 30 MHz to 26.5 GHz, and 30 MHz to 50 GHz.

FAQ 2012-02-26

Does the N5530S Measuring Receiver require a system calibration to meet specifications?
The N5530S does not require a system calibration. Its specifications are based upon the power meter and PSA performance.

FAQ 2012-02-26

What is the frequency range of the N5532B Sensor Module?
100 kHz to 4.2 GHz, 10 MHz to 18 GHz, 30 MHz to 26.5 GHz, and 30 MHz to 50 GHz.

FAQ 2012-02-26

Why do spurious signals appear in my spectrum analyzer span when I apply a clean signal to the RF input?
Overdriving the input mixer of the analyzer may cause the spurs.

FAQ 2012-02-26

When the API COM is closed on the N5530S system, is the calibration data cleared?
Yes. If you open the API COM connection, you should re-calibrate the system.

FAQ 2012-02-26

If we select another sensor module on the N5530S system, should we calibrate it each time?
Yes, you should calibrate the system when a sensor module is added.

FAQ 2012-02-26

What is the correct cable to use with the ESA Series for RS-232 communication?
Part number RS232-61601

FAQ 2012-02-25

Can I use 11970 or 11974 Series external mixers with my Keysight 859x Series Spectrum Analyzer?
No, while the 859X products did offer an L.O. output on the rear panel with option 010 Tracking Generator, there was never a provision for an I.F. Input, which is required for external mixing.

FAQ 2012-02-25

How does the 89400 series vector signal analyzer (VSA) lock to the symbols when it demodulates a vector-modulated signal?
There is no coherent carrier or symbol clock timing signal required for demodulation. The actual demodulation algorithm is proprietary, but...

FAQ 2012-02-25

What is the relationship between the Error Vector Magnitude and the Signal to Noise Ratio for a 64-QAM signal?
EVM is a voltage ratio of the error signal to the peak state voltage and SNR is a ratio of the signal power to the error power. The equation relating them is...

FAQ 2012-02-25

How can I transfer trace data from the 89400 VSA to my PC?
The VSA always saves data in standard data format; the utility is run from the DOS prompt.

FAQ 2012-02-25

How can I clear a "bad disk" or "bad file name" error message when accessing the internal RAM memory on the 89441A or 89410A?
If a "bad disk" error message occurs when trying to catalog a disk, or a "bad file name" message occurs when trying to copy a disk file, the non-volatile RAM memory (NVRAM disk) in the 89410A or 89441A has become corrupted.

FAQ 2012-02-25

How do the 89400 and 89600 families of vector signal analyzers (VSA) perform carrier-locking on a digitally-modulated signal?
First, the center frequency must be set to the carrier frequency within ±3% of the symbol rate...

FAQ 2012-02-25

Can a high-impedance probe be connected to the 89600S vector signal analyzer (VSA) in the same way as the 89410A VSA?
Instructions for 89400/89600 series Vector Signal Analyzer setup when using high impedance probes.

FAQ 2012-02-25

How do I save my trace data in a comma-separated variable (.csv) file format on an ESA, PSA or EMC spectrum analyzer?
On the front panel of the analyzer, press [File]{Save} and select either the A: (floppy) or C: (internal) drive...

FAQ 2012-02-24

How do I create an arbitrary waveform that I can download into the 89400 vector signal analyzer (VSA) and playback from its arbitrary (arb) source?
Here ia method for generating an arbitrary waveform from a math application like MATLAB or MathCad for use with the Keysight 89400 family of vector signal analyzers.

FAQ 2012-02-24

What are the rackmount accessories for a Keysight X-Series Signal Analyzer?
There are many different rackmount accessories that are compatible with the X-Series...

FAQ 2011-08-17

What are the differences between an E7400A Series EMC Analyzer and an ESA Series Spectrum Analyzer with Option AYQ - EMI Detectors & FM Demod?
See table provided....

FAQ 2011-07-07

Which high impedance / differential probe can be used with the E5052A/B Signal Source Analyzer?
The 54006A 6 GHz Passive Divider Probe Kit, the 41800A Active Probe, U1818A/B Active Differential Probe, the 1134A InfiniimMax 7 GHz Probe, can be used with the E5052A/B Signal Source Analyzer.

FAQ 2011-06-13

How do I determine and set the sample rate of the vector signal analyzer (VSA) for my application?
See content for details...

FAQ 2010-11-22

What signal sources are support for use as an alignment source for the PSA / N9039A RF Preselector based system?
8648B / N5181A / N5182A / N5183A/ N5161A / N5162A / E4438C / E8257D

FAQ 2010-09-08

Why is the marker Count Accuracy on the on the ESA Series Spectrum Analyzer only specified when using resolution bandwidths >1KHz?
If more that one signal is present in the 1kHz span, the marker count will be incorrect.

FAQ 2010-09-08

Can I change the front panel RF connector type on my N9039A RF Preselector?
Yes. See FAQ content for details...

FAQ 2010-09-08

85869PC: How to Fix Window Scrolling Problems?
This behavior is most likely being caused by your PC's display settings...

FAQ 2010-07-02

85869PC: Blank Screen When I Start the EMI Measurement Program
See FAQ content for details...

FAQ 2010-06-14