I am controlling the 8648X remotely. I have already turned off the pulse modulation to AM modulation, but the display still indicates 'PULSE'. Is the 8648X outputting an AM or PM modulation?

The AM modulation is activated now. An example of a workaround to correct the annotation is by the program below:

AM:STAT ON
AM:STAT OFF
PULM:STAT ON
PULM:STAT OFF
OUTP:STAT OFF
FM:STAT ON
FM:STAT OFF
OUTP:STAT ON
AM:STAT ON